HomeBlogAudiometer Calibration Requirements Under OSHA 1910.95: What Employers Actually Need to Know
compliance

Audiometer Calibration Requirements Under OSHA 1910.95: What Employers Actually Need to Know

Julia Johnson, Growth Lead, Soundtrace at SoundtraceJulia JohnsonGrowth Lead, Soundtrace10 min readApril 1, 2026
Audiometric Testing·OSHA Compliance·10 min read·Updated April 2026

Audiometer calibration is not a detail — it is the foundation of audiometric record validity. An audiometric test conducted with an out-of-calibration audiometer does not produce a valid OSHA record, regardless of how compliant every other aspect of the program is. OSHA 29 CFR 1910.95(h)(5) requires audiometers to be calibrated at least annually and checked acoustically before each day of testing. Audiometer calibration records that cannot be produced during an OSHA inspection or in WC litigation undermine the entire audiometric program. According to CDC/NIOSH, approximately 22 million U.S. workers undergo occupational audiometric testing annually.

OSHA Calibration Requirements Under Appendix E

OSHA 1910.95 Appendix E establishes the calibration requirements for audiometers used in HCP audiometric testing:

Calibration TypeFrequency RequiredWho PerformsWhat It Verifies
Biologic checkBefore each day of useTechnician (tests a person with known thresholds)Audiometer is producing audible tones; no obvious malfunction
Exhaustive calibrationAt least annuallyQualified technician with calibrated equipmentElectroacoustic output levels, frequencies, harmonic distortion, all per ANSI S3.6
Check after repairsWhenever audiometer is repairedQualified technicianFull ANSI S3.6 verification after any repair that could affect calibration
A Lapsed Calibration Invalidates the Records

An audiometer that has not been calibrated within the required annual window is out of compliance, and any audiometric records generated during the uncalibrated period may be challenged as unreliable. In OSHA inspections, missing calibration records result in citations. In WC proceedings, opposing counsel will argue that uncalibrated audiometric results should not be admitted or given weight. Calibration records must be maintained for the life of the audiometer program.

What the Daily Biologic Check Is and Isn’t

The daily biologic check is a functional verification, not a calibration. It confirms that the audiometer is producing audible tones and that the testing person with known thresholds obtains expected results. It does not verify electroacoustic output levels or frequency accuracy to ANSI S3.6 standards. A biologic check can catch obvious malfunctions but will not detect gradual calibration drift in output levels.

Both the daily biologic check and the annual electroacoustic calibration are required. Neither substitutes for the other.

Calibration Record Management

Maintain calibration records for every audiometer used in the program, including the date, technician, calibration results, and any adjustments made. Calibration records should be retained with audiometric program records — they are part of the documentation set that validates the audiometric results. In WC proceedings and OSHA inspections, the calibration record for the audiometer that generated a specific worker’s results may be specifically requested.


Frequently Asked Questions

What are OSHA’s audiometer calibration requirements under 1910.95?
OSHA 1910.95(h)(5) requires audiometers to be calibrated per Appendix E at least annually and checked acoustically before each day of use. Audiometers must meet ANSI S3.6 calibration standards. Calibration records must be maintained.
What is the difference between a biologic check and a full calibration?
A biologic check is a daily functional verification using a person with known stable thresholds. Full calibration to ANSI S3.6 requires electroacoustic measurement of output levels, frequencies, and distortion by a qualified technician with calibrated equipment, at least annually. Both are required; neither substitutes for the other.
What happens if audiometer calibration lapses?
Audiometric results from an uncalibrated audiometer may be challenged in OSHA inspections and WC litigation as unreliable. Missing calibration records result in OSHA citations and undermine the validity of all audiometric records generated during the uncalibrated period.

Calibration Maintained. Records Documented. Program Defensible.

Soundtrace audiometers maintain ANSI S3.6 calibration with documented records — supporting OSHA compliance and the audiometric record validity required for WC defense.

Get a Free Quote
Julia Johnson, Growth Lead, Soundtrace at Soundtrace

Julia Johnson

Growth Lead, Soundtrace, Soundtrace

Julia Johnson is the Growth Lead at Soundtrace, where she translates complex occupational health topics into clear, actionable content for safety professionals and employers. She works closely with the team to surface the insights and industry developments that matter most to hearing conservation programs.

Related Articles

Stay in the loop

Get compliance updates, product news, and practical tips delivered to your inbox.